Operation
|
Analytical principal: Anodic & Cathodic stripping voltammetry using unique vibrating microwire electrodes (patent appl No: PCT/GB2007/001704).
Parameters measured: Copper, Cadmium, Lea d, Zinc, Mercury and Arsenic
Temperature range: -20º to +70º
|
Data
|
Results obtained in less than 5 minutes.
Internal memory: 1000 data sets with facility to enter sample number, date and time stamp.
Transfer via USB to PC (optional Bluetooth). No specific software required.
New application methods can be uploaded to the unit via the USB port.
|
User Interface
|
LCD full graphics backlit display: 128mm x 128mm
Joystick cursor control
Menu driven software
|
Power
|
Rechargeable battery providing in excess of 50 tests per charge
Alternative power supply via mains adapter, vehicle ciga rette lighter or solar
power module option
|
Approvals
|
Waterproof to IP67
CE mark
|
Kit
|
Dimensions (mm): 470 (L) x 3 70 (W) x 170 (D)
Net weight: 9kg
|
Parameter
|
Lower
Limit
(ppb)
|
Upper
Limit
(ppb)
|
Accuracy
(ppB)
|
WHO
Guideline
Value (ppb)
|
Arsenic (As)
|
< 1.0
|
240.0
|
± 0.1
|
< 10.0
|
Copper (Cu)
|
0.1
|
240.0
|
± 0.01
|
< 2000.0
|
Mercury (Hg)
|
0.5
|
240.0
|
± 0.05
|
< 6.0
|
Zinc (Zn)
|
10.0
|
2200.0
|
±1.00
|
N/A
|
Cadmium (Cd)
|
0.5
|
2700.0
|
± 0.05
|
3.0
|
Lead (Pb)
|
0.5
|
2700.0
|
± 0.05
|
10.0
|